Skip Ribbon Commands
Skip to main content

DMT132 露点変換器 冷却式ドライヤー用

Vaisala HUMICAP® Dewpoint Transmitter DMT132 for refrigerant dryers
    カタログ

     技術仕様

    Dewpoint measurement


    ​Measurement range ​-20 ... +50 °C (-4 ... +122 °F) Td

    ​Accuracy at (+20 °C (+68 °F) ​±1 °C for -3...20 °C (+26.6 ... +68 °F) Td
    ±2 °C for -15...-3 °C (+5 ... +26.6 °F) Td

     

    Sensor

    Vaisala HUMICAP®180R

    Operating environment


    ​Operating temperature
    Operating pressure
    Relative humidity
    Sample flow rate
    Measured gases
    ​-10 ... + 60 °C (+14 ... +140 °F)
    0 ... 20 bar
    0 ... 100 %RH
    no effect on measurement accuracy
    non-corrosive gases

    Outputs


    ​Analog output (scalable)
    Resolution for current output
    Accuracy of analog outputs at
    +20 °C
    ​4...20 mA, 2-wire
    0.002 mA
    ±0.05% full scale

    ​Typical temperature
    dependence
    Connector

    ±0.005% of full scale/ °C
    4-pin M8( (IEC 60947-5-2)​

    General


    Recommended calibration interval
    (in refrigerant dryer application)
    Mechanical connection
    Operating voltage
    External load
    2 years


    G 1/2" ISO
    10 .... 28 VDC
    max 100 ohm for supply voltages <20 VDC
    max 500 ohm for supply voltages 20...28 VDC

    Weight
    Housing material
    Housing classification
    Storage temperature range
    Start-up time
    65 g (2.3 oz)
    PPS + 40% GF
    IP65 (NEMA 4)
    -40 ... + 80 °C (-40 ... +176 °F)
    3 s

    Options and accessories

     

    Tube filter
    Special cover set for HMK15
    NPT Adapter
    Sample cells

    230602
    230914
    210662SP
    DMT242SC, DMT242SC2, DSC74, DSC74B,
    DSC74C, DMCOIL

    For complete specifications, please see the brochure or manual in "Documents"

     
     
    サービス